* Not actual product image

00802

FMD-10K-L2 Force Measurement System 00802

The Starrett FMD-10K-L2 is a dual-column force measurement system optimized for high-volume production and quality control testing. The FMD-10K-L2 frame has a 10kN (2250lbf) testing capacity ideal for tensile or compressive testing. The system includes the test frame, controller & software. The controller features a Windows operating system with high resolution color, multi-touch display, Wi-Fi, Bluetooth and USB 2.0 port(s). The L2 software features test templates that require minimal operator setup in order to perform routine test methods such as push/pull to limit, push/pull to break and for determining peak results. For more complex, multi-stage test methods or methods that must comply with international testing standards, the L2 software features our exclusive Test Builder application.


Contact Starrett

Features

  • Turnkey 10kN (2250lbf) Force Measurement System includes Testing Frame, L2 Controller with L2 Force Measurement Software
  • Easily Perform Tension, Compression, Flexural, Shear, Peel and COF Testing
  • Perform Basic Force Test Methods Quickly with Easy-to-Use Test Templates
  • Perform Complex Multi-stage Testing Using Advanced Test Builder Application
  • Compact Footprint (31"W x 29"D)

Specifications

  • Frame Type: Dual Column
  • User Interface: Tablet PC
  • Load Capacity: 10kN, 2250lbf, 1000kgf
  • Maximum Speed: 60"/min, 1525mm/min
  • Minimum Speed: 0.00004"/min, 0.001mm/min
  • Maximum Load @ Full Speed: 10kN, 2250lbf, 1000kgf
  • Maximum Speed @ Full Load: 60"/min, 1525mm/min
  • Return Speed: 60"/min, 1525mm/min
  • Total Crosshead Travel: 45.75", 1162mm
  • Total Vertical Test Space: 50", 1270mm
  • Crosshead Throat Depth: 16.7", 424mm
  • Height: 64", 1626mm
  • Width: 31", 787mm
  • Depth: 29", 736mm
  • Weight: 400lbs, 181kg
  • Accuracy - Load Measurement: Load Cell Dependant
  • Accuracy - Strain Measurement: N/A
  • Load Cell Sensor Models: Models FLC Series, MLC Series, or ULC Series
  • Extensometer I/O: I/O Not Availible
  • USB I/O: I/O Standard
  • Power: Single Phase Voltage (Vac) +/-10% 100, 120, 200, or 240
  • Operating Temperature: 50 to 100F, (10 to 38F)
  • Humidity Range: 10 to 90% non-condensing
  • CE Compliant: Yes